Product characteristics
| Number of inputs and outputs | Number of digital outputs: 2; Number of analogue outputs: 1 |
|---|---|
| Process connection | G 1/4 DN6 flat seal |
Application
| Special feature | Gold-plated contacts | ||
|---|---|---|---|
| Media | conductive liquids; water; hydrous media | ||
| Note on media |
|
||
| Medium temperature [°C] | -20…90 | ||
| Pressure rating [bar] | 16 | ||
| Pressure rating [Mpa] | 1.6 |
Electrical data
| Operating voltage [V] | 18…30 DC; (to SELV/PELV) |
|---|---|
| Current consumption [mA] | < 80 |
| Protection class | III |
| Reverse polarity protection | yes |
| Power-on delay time [s] | 5 |
Inputs / outputs
| Number of inputs and outputs | Number of digital outputs: 2; Number of analogue outputs: 1 |
|---|
Inputs
| Inputs | counter reset |
|---|
Outputs
| Total number of outputs | 2 |
|---|---|
| Output signal | switching signal; analogue signal; pulse signal; IO-Link; frequency signal; (configurable) |
| Electrical design | PNP/NPN |
| Number of digital outputs | 2 |
| Output function | normally open / normally closed; (parameterisable) |
| Max. voltage drop switching output DC [V] | 2 |
| Permanent current rating of switching output DC [mA] | 100 |
| Number of analogue outputs | 1 |
| Analogue current output [mA] | 4…20; (scalable) |
| Max. load [Ω] | 500 |
| Pulse output | flow rate meter |
| Short-circuit protection | yes |
| Type of short-circuit protection | pulsed |
| Overload protection | yes |
Measuring/setting range
| Measuring range [ml/min] | 5…5000 |
|---|---|
| Display range [ml/min] | -6000…6000 |
| Resolution [ml/min] | 1 |
| Set point SP [ml/min] | 33…5000 |
| Reset point rP [ml/min] | 7…4974 |
| Analogue start point ASP [ml/min] | 0…3993 |
| Analogue end point AEP [ml/min] | 1007…5000 |
| Low flow cut-off LFC [ml/min] | 5…250 |
| Frequency end point, FEP [ml/min] | 1005…5000 |
| Frequency at the end point FRP [Hz] | 1…10000 |
Volumetric flow quantity monitoring
| Pulse length [s] | 0.005…2 |
|---|---|
| Pulse value | 0.001…99990000 I |
Temperature monitoring
| Measuring range [°C] | -20…90 |
|---|---|
| Display range [°C] | -42…112 |
| Resolution [°C] | 0.1 |
| Set point SP [°C] | -19.6…90 |
| Reset point rP [°C] | -20…89.6 |
| Analogue start point [°C] | -20…68 |
| Analogue end point [°C] | 2…90 |
| In steps of [°C] | 0.1 |
Accuracy / deviations
Flow monitoring
| Accuracy (in the measuring range) | ± (0,8 % MW + 0,2 % MEW) |
|---|---|
| Repeatability | ± 0,2 % MEW |
Temperature monitoring
| Accuracy [K] | ± 2,5 (Q > 1 l/min) |
|---|
Response times
Flow monitoring
| Start-up delay [s] | 0…50 |
|---|---|
| Response time [s] | < 0.25; (dAP = 0, T09) |
| Damping process value dAP [s] | 0…5 |
Temperature monitoring
| Response time [s] | 15; (Q > 1 l/min, T09) |
|---|
Software / programming
| Parameter setting options | hysteresis / window; normally open / normally closed; switching logic; frequency output; current/pulse output; start-up delay; display can be deactivated; Display unit |
|---|
Interfaces
| Communication interface | IO-Link | ||||
|---|---|---|---|---|---|
| Transmission type | COM2 (38,4 kBaud) | ||||
| IO-Link revision | 1.1 | ||||
| SDCI standard | IEC 61131-9 | ||||
| Profiles | Smart Sensor: Process Data Variable; Device Identification, Device Diagnosis | ||||
| SIO mode | yes | ||||
| Required master port type | A | ||||
| Process data analogue | 3 | ||||
| Process data binary | 2 | ||||
| Min. process cycle time [ms] | 6 | ||||
| Supported DeviceIDs |
|
Operating conditions
| Ambient temperature [°C] | -20…60 |
|---|---|
| Storage temperature [°C] | -25…80 |
| Protection | IP 65; IP 67 |
Tests / approvals
| EMC |
|
||||
|---|---|---|---|---|---|
| Shock resistance |
|
||||
| Vibration resistance |
|
||||
| MTTF [years] | 114 | ||||
| UL approval |
|
||||
| Pressure Equipment Directive | Sound engineering practice; can be used for group 2 fluids; group 1 fluids on request |
Mechanical data
| Weight [g] | 706.9 |
|---|---|
| Materials | stainless steel (1.4408/316); stainless steel (1.4404 / 316L); PC; PBT+PC-GF30 |
| Materials (wetted parts) | stainless steel (1.4404 / 316L); PEEK; carbon fibre PEEK; EPDM; Centellen |
| Process connection | G 1/4 DN6 flat seal |
Displays / operating elements
| Display |
|
|---|
Remarks
| Remarks |
|
||
|---|---|---|---|
| Pack quantity | 1 pcs. |
Electrical connection
| Connection | Connector: 1 x M12; coding: A; Contacts: gold-plated |
|---|








Reviews